National Stock Number: 6665-01-515-7777
Federal Supply Class: 6665
National Item Identification Number: 015157777
Description: DETECTOR,RADIAC
Detail: A DEVICE CONSISTING OF A PERMANENT OR REPLACEABLE ELEMENT WHICH IS SENSITIVE TO RADIOACTIVITY OR FREE NUCLEAR PARTICLES, TOGETHER WITH ITS ASSOCIATED CIRCUITS AND HOUSING. AS A RESULT OF THE INCIDENT IONIZING RADIATION, IT PRODUCES AN ELECTRON PULSE OR CURRENT FLOW THAT CAN BE INTERPETED BY OTHER COMPONENTS. DOES NOT INCLUDE METERS OR INDICATORS. EXCLUDES PROBE, RADIAC.
|
Manufacturer Information:
RADCAL 10X5-180 | 4P061 | RADCAL CORP |
|
Techincal Specification:
FUNCTIONAL DESCRIPTION | DESIGNED FOR LEAKAGE MEASUREMENTSAS DESCRIBED IN REGULATIONS FOR THE ADDMINISTRATION AND ENFORCEMENT OF THE RADIATION CONTROL FOR HEALTH AND SAFETYACT OF 1968 AND PROVIDES THE REQUIRED EFFECTIVE CROSS-SECTION OF 100 CM2 | SPECIAL FEATURES | CHAMBER OVERVIEW-ROENTGENS RATE RESOLUTION 0.1 MR/HR; MAX 1.7 KR/HR; DOSE RESOLUTION 0.01 UR; MAX 2.3 KR; RATE SPECIFICATIONS 0.1 MR/HR-1.7 R/HR; 1UGY/HR-15 GY/HR; EXPOSURE SPECIFICATIONS 0.01 UR-2.3 KR; 0.1 NGY-20 GY; CALIBRATION ACCURACY PORM 4 PCT USING X-RAYS AT 150 KVP AND 10.2 MM AI HVL; EXPOSURE RATE DEPENDENCE PORM 5 PCT, 20 MR/HR TO 2000 R/HR; ENERGY DEPENDENCE PORM 5 PCT, 33 KEV TO 1.33 MEV, WITH BUILD-UP MATERIAL; CONSTRUCTION POLYCARBONATE WALLS AND ELECTRODE; CONDUCTING GRAPHITE EXTERIOR COATING; 180 CM3 ACTIVE VOLUME; 0.11 KG | SUPPLEMENTARY FEATURES | CONTROLLING AGENCY NOMENCLATURE LEAKAGE AND LOW LEVEL MEASUREMENT CHAMBER |
|
|