ANODE HEAT CAPACITY RATING IN HEAT UNITS | 300000 |
CONTAINER TYPE | ANY ACCEPTABLE |
FOCAL SPOT QUANTITY | 2 |
FOCAL SPOT SIZE | 0.6 MM SQ; 1.5 MM SQ |
HOUSING X-RAY PROTECTIVE VALUE | X-RAY PROTECTIVE VALUE OF HOUSING SHALL BE IN ACCORDANCE WITH DHEW PUBLICATION NO. FDA 75-8003 JUL 1974 |
PART NAME ASSIGNED BY CONTROLLING AGENCY | TUBE UNIT MEDICAL X-RAY APPARATUS FOCAL SPOT SIZE 0.6;1.5MM SQ |
PORT TO CABLE RECEPTACLE ANGLE IN DEG | 90.0 |
TUBE INSERT RADIOGRAPHIC PEAK VOLTAGE RATING | 150.0 KILOVOLTS |
TUBE INSERT ANODE TYPE | HIGH SPEED ROTATING |
TARGET ANGLE IN DEG | 15.0 |
TARGET BASE MATERIAL | MOLYBDENUM |
TARGET FACE MATERIAL | TUNGSTEN-RHENIUM ALLOY |